Open original FDA record ↗Recall record
- FDA identifier
- Z-1812-2025
- Company in FDA record
- Copan WASP
- FDA classification date
- May 21, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.
Source facts
- Action
- On 5/6/2025, recall notices were mailed to customers informing them of the following: Ensure the plate images are reviewed against the actual agar plate to verify the assessment you performed by the digital image inspection. Complete and return the response form via email to regulatory.wasp@copangroup.com
- Initiated
- 2025-05-06
- Root cause
- No Marketing Application
- Distribution
- US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.
- Recall number
- Z-1812-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213566
- FDA classified
- 2025-05-21
- Recall event number
- 96765