Skip to content

Public FDA record

Z-1813-2025: HiResolution" Bionic Ear System - M Battery Charger Model Number CI-5607

Recall record Z-1813-2025 for Advanced Bionics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1813-2025
Company in FDA record
Advanced Bionics, LLC
FDA classification date
May 22, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect the battery charger being packaged and distributed. Product labeling does not match the intended product as specified.

Source facts

Action
On 03/06/2025, customer service informed customers via telephone call of a distribution error. An updated "URGENT: Medical Device Recall Notification will be communicated to customers on or about 04/09/2025, via email. This updated notification will inform them that upon further review/investigation, it was identified that the individual package label was for the Marvel M Battery Charger (CI-5607), however the product within the package was a Naida CI PowerCel Charger (CI-5605). Customer are informed that: 1. The mislabeled battery charger they received was from lot either Lot Numbers: 410488967 and/or 410488966 Customers are instructed to: 2. Return the affected chargers to Advance Bionics. Advance Bionics will provide the correct charger to customer and will confirm receipt and return of the affected chargers. For questions contact customer service at 877-829-0026 or email Customerservice@advancebionics.com
Initiated
2025-03-05
Root cause
Under Investigation by firm
Distribution
U.S.: AZ, FL, MI, SC, TN and TX. O.U.S.: N/A
Recall number
Z-1813-2025
Recall status
Open, Classified
CDRH record ID
213379
FDA classified
2025-05-22
Recall event number
96668

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗