Open original FDA record ↗Recall record
- FDA identifier
- Z-1814-2025
- Company in FDA record
- SunMed Holdings, LLC
- FDA classification date
- May 30, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.
Source facts
- Action
- Recall initiated May 1, 2025. Customers instructed to immediately cease use of any affected product and quarantine all affected product. Customers were instructed to return or destroy affected product. AirLife will send new replacement product to customers once disposition is confirmed. If you need replacement products to be sent to you urgently, please call AirLife directly on 1-800-433-2797 and the firm will make every effort to accommodate your needs.
- Initiated
- 2025-05-01
- Root cause
- Process control
- Distribution
- US Nationwide distribution via Medline.
- Recall number
- Z-1814-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213696
- FDA classified
- 2025-05-30
- Recall event number
- 96801