During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
Source facts
Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to
all affected customers dated March 9, 2026.
Initiated
2026-03-12
Root cause
Radiation Control for Health and Safety Act
Distribution
U.S. and O.U.S.
Recall number
Z-1814-2026
Recall status
Open, Classified
CDRH record ID
219758
FDA classified
2026-04-16
Recall event number
98772