Open original FDA record ↗Recall record
- FDA identifier
- Z-1815-2012
- Company in FDA record
- Tripath Imaging, Inc.
- FDA classification date
- Jun 15, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class III
Reason stated in the source
The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
Source facts
- Action
- BD Diagnostics sent a "BD PrepMate Syringing Pipette" letter dated March 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm encouraged customers to refer to the Operator's Manual and inspect the centrifuge tubes to verify that the correct amount of sample has been transferred. Customers experiencing this issue were instructed to discard the pipettes and contact BD Diagnostic Technical Support for a replacement. Contact the firm at 1-877-822-7771 for questions regarding this notice.
- Initiated
- 2012-03-22
- Root cause
- Component design/selection
- Terminated
- 2012-12-03
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1815-2012
- Recall status
- Terminated
- CDRH record ID
- 108863
- FDA classified
- 2012-06-15
- Recall event number
- 61737