Z-1815-2022: Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An

Recall record Z-1815-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1815-2022
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Sep 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment

Source facts

Action
Olympus issued Urgent Medical Device Recall Letter on 8/12/22 to Surgical Department Risk Management. Letter states reason for recall, health risk and action to take: 1. Immediately locate any product with affected lot numbers listed in this communication. See Attachment 2 for pictures showing the location of lot numbers on the package. 2. Call your Olympus customer service representative at 1 800 848 9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number 0412 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this removal action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact directly at (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Initiated
2022-08-12
Root cause
Under Investigation by firm
Distribution
Nationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA
Recall number
Z-1815-2022
Recall status
Open, Classified
CDRH record ID
195727
FDA classified
2022-09-26
Recall event number
90788

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗