Recall record Z-1815-2025 for Beckman Coulter Ireland, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1815-2025
Company in FDA record
Beckman Coulter Ireland, Inc.
FDA classification date
May 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

Source facts

Action
On April 23, 2025, Beckman Coulter issued a "Urgent Medical Device Recall" Notification to affected consignees via mail and E-Mail. Beckman Coulter asked consignees to take the following actions: 1. If your laboratory is reporting Apo B results generated using the combination of reagent OSR6143, Lot 2631 and calibrator ODR3022, Lot 3000 in mg/dL units, a retrospective review of patient results greater than 77 mg/dL and less than 176 mg/dL is recommended. 2. Discard the value assignment sheet located within each kit of Apo B OSR6143 Lot 2631. 3 Download the updated VAS kit insert from the Beckman Coulter website by searching LOSR6143_2631_ML.02. Beckman Coulter Diagnostics | Beckman Coulter. 4. Ensure that the level 4 calibrator ODR3022, Lot 3000 target value used for calibration of Apo B OSR6143 Lot 2631 is 116 mg/dL and aligns with the corrected VAS. 5. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter.
Initiated
2025-04-23
Root cause
Error in labeling
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.
Recall number
Z-1815-2025
Recall status
Open, Classified
CDRH record ID
213512
FDA classified
2025-05-23
Recall event number
96746

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗