Skip to content

Public FDA record

Z-1816-2013: Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800

Recall record Z-1816-2013 for Synthes USA HQ, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1816-2013
Company in FDA record
Synthes USA HQ, Inc.
FDA classification date
Jul 29, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.

Source facts

Action
Synthes sent a Urgent Notice Medical Device Recall letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is requesting that you please examine your inventory for products with the above Part and Lot number and remove them. Call Synthes at 1-800-479-6329. If you have any further questions please call 610-719-5450.
Initiated
2012-10-12
Root cause
Packaging
Distribution
US Distribution including the states of CA, MN, NC and TN.
Recall number
Z-1816-2013
Recall status
Open, Classified
CDRH record ID
119840
FDA classified
2013-07-29
Recall event number
65669

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗