Open original FDA record ↗Recall record
- FDA identifier
- Z-1816-2013
- Company in FDA record
- Synthes USA HQ, Inc.
- FDA classification date
- Jul 29, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.
Source facts
- Action
- Synthes sent a Urgent Notice Medical Device Recall letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is requesting that you please examine your inventory for products with the above Part and Lot number and remove them. Call Synthes at 1-800-479-6329. If you have any further questions please call 610-719-5450.
- Initiated
- 2012-10-12
- Root cause
- Packaging
- Distribution
- US Distribution including the states of CA, MN, NC and TN.
- Recall number
- Z-1816-2013
- Recall status
- Open, Classified
- CDRH record ID
- 119840
- FDA classified
- 2013-07-29
- Recall event number
- 65669