Open original FDA record ↗Recall record
- FDA identifier
- Z-1817-2020
- Company in FDA record
- WELCH ALLYN, INC/MORTARA
- FDA classification date
- Apr 29, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Source facts
- Action
- Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed.
- Initiated
- 2020-03-18
- Root cause
- Device Design
- Terminated
- 2026-02-09
- Distribution
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
- Recall number
- Z-1817-2020
- Recall status
- Terminated
- CDRH record ID
- 180595
- FDA classified
- 2020-04-29
- Recall event number
- 85297