Z-1817-2024: OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1

Recall record Z-1817-2024 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1817-2024
Company in FDA record
Philips Respironics, Inc.
FDA classification date
May 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Device may experience an interruption or loss of therapy in case of a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death.

Source facts

Action
Firm notified customers through mailed letter "URGENT Medical Device Recall" on April 1, 2024. The letter described the product, problem and actions to be taken. The firm reminded customers to refer to the user manual and to perform a clinical assessment prior to placing a patient on the ventilator to ensure that: 1. The device is appropriately set for patient requirements. 2. Alternate ventilation equipment is available. 3. Alternative monitoring is used where appropriate. Depending on the level of ventilatory support required, customers were advised to place patients on an alternate source of ventilation. Philips is currently investigating the issue. If you need any further information or support concerning this issue, please contact your local Philips Respironics representative: 1-800-345-6443, prompts 4, 5 or email at respironics.clinical@philips.com. A second "URGENT Medical Device Recall" notice was distributed to customers on June 12, 2024. In the notice customers were provided additional guidance on how to operate the device when a Ventilator Inoperative Alarm occurs and how to perform an optional hard reboot of their device that will temporarily restore functionality. An UPDATE URGENT Medical Device Recall Notice was mailed to consignees on 5/20/25 providing instructions on how to perform a Hard Reboot in the instance of a Ventilator Inoperable Alarm condition. Consignees are instructed to have users of the device review the OLA+ User Manual and the provided Appendix A.
Initiated
2024-04-01
Root cause
Device Design
Distribution
US Nationwide distribution and OUS International distribution to Canada.
Recall number
Z-1817-2024
Recall status
Open, Classified
CDRH record ID
207300
FDA classified
2024-05-23
Recall event number
94487

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗