Open original FDA record ↗Recall record
- FDA identifier
- Z-1817-2025
- Company in FDA record
- Onkos Surgical, Inc.
- FDA classification date
- May 23, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.
Source facts
- Action
- ONKOS Surgical notified their distributors on about 12/22/2023 via email. They were instructed to segregate and quarantine the product immediately, complete and return the "Product and Distribution Information" form within 2 business days and return any existing implants on hand within 2 business days after signing the acknowledgement. They were also instructed that they will receive replacement implants and to place them in their modular segmental collar stem implant packs.
- Initiated
- 2023-12-22
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide distribution in the states of California, Kentucky, and Wisconsin.
- Recall number
- Z-1817-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213655
- FDA classified
- 2025-05-23
- Recall event number
- 96781