Z-1818-2011: Access Thyroglobulin Antibody II, Part Number: A32898 The Access Thyroglobulin Antibody II (TgAb) assay

Recall record Z-1818-2011 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1818-2011
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Mar 24, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated after Beckman Coulter confirmed customer reports of reagent pack to pack variability for the Access Thyroglobulin Antibody II (TgAb) reagent lot numbers identified in the recall. Erroneous results for patient and quality control samples can be generated. The Thyroglobulin Antibody II (TgAbII) reagent lots may produce: (1) Erroneously high TgAb results for both quality co

Source facts

Action
Beckman Coulter sent a Product Corrective Action (PCA) letter on January 26, 2011 via US Postal Service to the affected customers. The letter explained the problem identified and the action to be taken. Customers were instructed to: (1) Discontinue use of Access Thyroglobulin Antibody II reagent lot numbers identified in the table above; (2) If customers obtained questionable thyroglobulin antibody results, or received feedback from clinicians on reported thyroglobulin antibody results using the reagent lot numbers identified above, rerun or request a redraw of the patient sample; (3) Share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If they forwarded any of the affected product listed above to another laboratory, they should provide a copy of the letter to them; (4) Complete and return the enclosed response form within 10 days so the firm can be assured that you have received this important notification and are taking appropriate action. For questions regarding the recall communication, customers were told to contact Beckman Coulter Customer Support Center: (1) Via our website, http://www.beckmancoulter.com/customersupport/support (2) Via phone, call 1-800-854-3633 in the United States and Canada (3) Outside the United States and Canada please contact your local Beckman Coulter representative.
Initiated
2011-01-26
Root cause
Other
Terminated
2012-05-04
Distribution
Worldwide Distribution - USA including AR, CA, FL, GA, HI, IA, IL, KY, MA, MI, MN, MO, MS, NC, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI and the countries of Argentina, Armenia, Australia, Belgium, Brazil, Bulgaria, Canada, China, Czech Republic, France, French Polynesia, Georgia, Germany, Greece, Hungary, Israel, Italy, Kazakhstan, Kuwait, Lebanon, Morocco, Netherlands, New Zealand, Oman, Poland, Puerto Rico, Romania, Russian Federation, Serbia, South Africa, Spain, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Arab Emirates, Yemen
Recall number
Z-1818-2011
Recall status
Terminated
CDRH record ID
97800
FDA classified
2011-03-24
Recall event number
57946

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗