Z-1818-2012: Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationar

Recall record Z-1818-2012 for Philips Healthcare Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1818-2012
Company in FDA record
Philips Healthcare Inc.
FDA classification date
Jun 15, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side

Source facts

Action
Philips Healthcare issued a Field Safety Notice on 5/22/12 advising users of the error. A software update (Rel 2.1.2) will issue to correct the problem. Contact your local Philips representative. 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Initiated
2012-05-22
Root cause
Software change control
Terminated
2017-02-24
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Canada AFGHANISTAN ALBANIA ALGERIA ANGOLA ARGENTINA ARMENIA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BAHRAIN BANGLADESH BARBADOS BELARUS BELGIUM BENIN BHUTAN BOLIVIA BOSNIA&HERZEGOVINA BRAZIL BRUNEI DARUSSALAM BULGARIA BURUNDI CAMBODIA CAMEROON CANARY ISLANDS CHAD CHILE CHINA COLOMBIA CONGO CONGO,DEMOCRATIC COSTA RICA COTE D'IVOIRE CROATIA CUBA CYPRUS CZECH REPUBLIC DENMARK DJIBOUTI DOMINICAN REPUBLIC ECUADOR EGYPT EL SALVADOR EQUATORIAL GUINEA ERITREA ESTONIA ETHIOPIA fake test ssd FINLAND FRANCE FRENCH GUIANA GABON GAMBIA GEORGIE GERMANY GHANA GREECE GUADELOUPE GUAM GUATEMALA GUINEA GUYANA HONDURAS HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAMAICA JAPAN JORDAN KAZAKSTAN KENYA KUWAIT KYRGYZSTAN LAO PEOPLE'S DEM.REP LATVIA LEBANON LIBERIA LIBYAN ARAB JAMAHIRI LIECHTENSTEIN LITHUANIA LUXEMBOURG MACAU MACEDONIA MALAWI MALAYSIA MALI MALTA MARTINIQUE MAURITANIA MAURITIUS METRACOM MEXICO MOLDOVA MONGOLIA MOROCCO MOZAMBIQUE NAMIBIA NEPAL NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NICARAGUA NIGERIA NORTH KOREA NORWAY OMAN PAKISTAN PALESTINE PANAMA PARAGUAY PERU PHC BEST PHILIPPINES PMCC BU Canada PMCC BU USA POLAND PORTUGAL PUERTO RICO QATAR REUNION ROMANIA RUSSIA RWANDA SAUDI ARABIA SENEGAL SERBIA&MONTE NEGRO SIERRA LEONE SINGAPORE SLOVAKIA SLOVENIA SOMALIA SOUTH AFRICA SOUTH KOREA SPAIN SRI LANKA SUDAN SURINAME SWEDEN SWITZERLAND SYRIAN ARAB REPUBLIC TAIWAN TAJIKISTAN TANZANIA THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY TURKMENISTAN UGANDA UKRAINE Unidentified Country UNITED ARAB EMIRATES UNITED KINGDOM URUGUAY UZBEKISTAN VENEZUELA VIET NAM YEMEN ZAMBIA ZIMBABWE
Recall number
Z-1818-2012
Recall status
Terminated
CDRH record ID
109499
FDA classified
2012-06-15
Recall event number
61926

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗