Skip to content

Public FDA record

Z-1819-2008: Allez Spine Laguna Pedicle Screw System: IT-FN1001, Laguna Sterilization Tray # 1 (Part Number: 54003);

Recall record Z-1819-2008 for Allez Spine, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1819-2008
Company in FDA record
Allez Spine, LLC
FDA classification date
Sep 2, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This field action was initiated because the cleaning, decontamination and sterilization procedures specified in the "Instructions for Use" for the impacted systems are incorrect. The use of these parameters in performing cleaning, decontamination and sterilization procedures may not result in clean and sterile product.

Source facts

Action
Allez Spine issued a safety notice with attached fax-back attachment on April 14, 2008 via FedEx to all its hospital users, distributors and field representatives informing them of the failure of the sterilization validation and requesting them to use the new recommended sterilization parameters. The notice included copies of the updated Instructions for Use for the Allez Spine Laguna Pedicle Screw System and the Allez Spine Del Mar Pedicle Screw System. The safety notice requested that the revised recommended instructions be used for all future cleaning, decontamination and sterilization. Copies of the revised instructions were also provided with the safety notice along with links to the company website where additional copies may be downloaded. Contact Mark De Baca at 415-357-9908 if you have any questions.
Initiated
2008-04-14
Root cause
Error in labeling
Terminated
2010-01-25
Distribution
Nationwide to AZ, CA, CO, GA, IL, NV, TN, TX, UT
Recall number
Z-1819-2008
Recall status
Terminated
CDRH record ID
70846
FDA classified
2008-09-02
Recall event number
48116

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗