Open original FDA record ↗Recall record
- FDA identifier
- Z-1819-2015
- Company in FDA record
- Orthofix, Inc
- FDA classification date
- Jun 19, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.
Source facts
- Action
- Orthofix sent an Urgent Medical Device Product Advisory Notification letter dated May 11, 2015, to the affected distributors and hospitals. The consignees were directed to inspect thier inventory and perform checks before surgical use of the drill guides. Customers were instructed to provide a copy of the Product Advisory Notification letter if product was further distributed. Customers with questions were instructed to contact their local Orthofix representative and to for returns they should contact Orthofix customer service representative at 888-298-5700.
- Initiated
- 2015-05-11
- Root cause
- Device Design
- Terminated
- 2016-02-17
- Distribution
- Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.
- Recall number
- Z-1819-2015
- Recall status
- Terminated
- CDRH record ID
- 137146
- FDA classified
- 2015-06-19
- Recall event number
- 71262