Open original FDA record ↗Recall record
- FDA identifier
- Z-1819-2017
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Apr 13, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
Source facts
- Action
- Firm sent e-mails on 3-17-17 to affected sites instructing them to STOP USE of the XI Suction Irrigator and RMA. and return all unused Xi Suction/irrigation instruments for credit. A formal letters was sent to affected sites on March 31, 2017. Letters advised of the reason for recall with details on product name and product number. Letters requested that any unused devices be returned to Intuitive via the standard RMA process. All personnel who use the Da Vinci system should be made aware of the issue. Response form should be completed online or by returning copy enclosed with the letter.
- Initiated
- 2017-03-31
- Root cause
- Software design
- Terminated
- 2017-07-28
- Distribution
- US Only - one location each in AL, CO, KS, NV, NY, and 2 in TX
- Recall number
- Z-1819-2017
- Recall status
- Terminated
- CDRH record ID
- 154548
- FDA classified
- 2017-04-13
- Recall event number
- 76911