OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Source facts
Action
All impacted customers with receive a customer letter notifying them of the potential
issue. GE HealthCare will send a field service engineer to perform measurements of the Xray
field on the identified systems (listed in Appendix A of the customer letter).
Initiated
2025-03-21
Root cause
Process control
Distribution
US and OUS
Recall number
Z-1819-2025
Recall status
Open, Classified
CDRH record ID
214001
FDA classified
2025-06-09
Recall event number
96902