Z-1820-2022: IMMULITE 2000 and IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) Siemens Material Number (SM

Recall record Z-1820-2022 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1820-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Sep 27, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid

Source facts

Action
Siemens Healthcare issued an "Urgent Medical Device Correction (UMDC IMC 22-09.A.US)" to US customers, and an "Urgent Field Safety Notice (UFSN IMC 22-09.A.OUS)" to Outside the US (OUS) customers. The UMDC and UFSN letters provide instructions to customers to follow: When this issue occurs, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid. Mitigations would include correlation of tests results with additional tests such as anti-TPO antibodies along with thyroid function tests such as total T3 and T4, free T3 and T4 and TSH results. A review of previously generated results is not recommended as results would not be used in isolation. Actions to be Taken by the Customer: Please review this letter with your Medical Director. Discontinue use of and discard the Kit Lot listed in Table 1. Review your inventory of this product to determine your laboratory's replacement needs and to provide information to Siemens for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days via email to: uscctsfcaecfax.team@siemens-healthineers.com Or to fax this completed form to the Customer Care Center at: (312) 275-7795 If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, contact your local Siemens Healthineers technical support representative or call 508-668-5000.
Initiated
2022-08-11
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US (nationwide) to states of: FL IN MA MI NH NJ NY OH TX UT and OUS (Foreign) countries of: Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Brazil¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ P.R.¿China¿¿¿¿¿ Singapore¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿
Recall number
Z-1820-2022
Recall status
Open, Classified
CDRH record ID
195933
FDA classified
2022-09-27
Recall event number
90812

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗