Open original FDA record ↗Recall record
- FDA identifier
- Z-1821-2015
- Company in FDA record
- Unomedical As
- FDA classification date
- Jun 19, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unomedical has found that in rare cases the steel needle can break during use, interrupting the delivery of medication including insulin.
Source facts
- Action
- Each consignee was sent an "Urgent Field Safety Notice Regarding Steel Cannula Infusion Sets" on 5/21/2015 via Express Mail. This notification identified the affected product as well as the reason for the recall. It also contained instructions on how to use the product. Customers were to contact their health care provider if they suspected that a needle had broken off and remained under the skin. Diabetic patients were to monitor their blood sugar and non diabetic patients were to monitor symptoms which would indicate drug delivery had failed. Questions should be directed to the persons identified in the letter.
- Initiated
- 2015-05-21
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-01-23
- Distribution
- Worldwide Distribution -- US, including the states of CA, IL, MN, and PA; and, the countries of Australia, Bahrain, Belgium, Canada, China, Cypress, Denmark, Estonia, France, Germany, Greece, Hong Kong, Hungary, Israel, Italy, Kuwait, Lebanon, Malaysia, Maldives, Oman, Schweiz, South Africa, Spain, Sweden, Switzerland, Netherlands, Turkey, United Arab Emirates, and United Kingdom.
- Recall number
- Z-1821-2015
- Recall status
- Terminated
- CDRH record ID
- 137324
- FDA classified
- 2015-06-19
- Recall event number
- 71332