Z-1821-2022: OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA01001L OSSUR Power

Recall record Z-1821-2022 for Ossur Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1821-2022
Company in FDA record
Ossur Americas
FDA classification date
Sep 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Battery may dislodge from the Power Knee

Source facts

Action
On 08/29/2022, The firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that reports have been received of units where the battery dislodged from the Power Knee units, causing loss of power to the device. This may cause an increase risk for stumbles or falls, potentially leading to an injury. Customer are being instructed to implement a correction to retrofit a security tab to their units. The retrofit kit will be provided with clear instructions for attaching the security tab to the device. In addition, customer are asked to: 1. Pass this notice to those who need to be aware within their organization or to any organization where Power Knee devices have been transferred. 2. Reconcile Ossur s records with their own to ensure all devices are accounted. 3. Once all users have been identified and the retrofit kits have been received, customers will need to schedule each individual to come into their clinic to complete the retrofit process. 4. Sign the acknowledgment form confirming receipt of this notice and that the retrofit process has been completed for all affected devices. Ossur is here to support customers through the process. Contact customer service for further information and assistance. Tel: 1-800-233-6263.
Initiated
2022-08-29
Root cause
Device Design
Distribution
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, VA, VT, WA and WI.
Recall number
Z-1821-2022
Recall status
Open, Classified
CDRH record ID
195707
FDA classified
2022-09-28
Recall event number
90775

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗