Open original FDA record ↗Recall record
- FDA identifier
- Z-1822-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 25, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Loss of image data: Software anomaly in the Processing Software on the Definium 8000 system may impact patient safety when using the VolumeRAD advanced application (option). The slice visualization of VolumeRAD exams acquired on the Wallstand receptor will be offset by 17mm. This prevents the visualization of anatomy WITHIN a distance of 17mm from the wallstand patient barrier (receptor cover).
Source facts
- Action
- GE sent a Product Safety Notification, dated March 7, 2008, to all customers. The letter described the safety issue, products affected, and stated that a GE Healthcare Field Engineer will schedule a visit to each facility to install a software modification.
- Initiated
- 2008-03-07
- Root cause
- Software design
- Terminated
- 2011-12-11
- Distribution
- Worldwide distribution: USA, Australia, Belgium, Canada, China, Finland, France, Germany, Great Britain, Italy, Netherlands, Spain, and Sweden.
- Recall number
- Z-1822-2008
- Recall status
- Terminated
- CDRH record ID
- 73650
- FDA classified
- 2008-09-25
- Recall event number
- 49504