Open original FDA record ↗Recall record
- FDA identifier
- Z-1823-2011
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Mar 25, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm has recieved complaints of imprecision and erroneous patient results for calcium on the COBAS INTEGRA and COBAS c 501 system using reagent lots 62601901 and 63020401.
Source facts
- Action
- The firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated October 11, 2010 to all consignees/customers via UPS(signature required) . The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the recalled product; discard from inventory according to their site's local regulations; complete and return the attached Faxback form via fax to 1-877-598-5705 even if they do not have the affected product, and file this letter for future reference. The firm added lot 63020401 to the recall in 3/2011. The firm is mailing an URGENT MEDICAL DEVICE REMOVAL notice to customers that purchased that lot of product. Please contact Roche Diagnostics Technical Support at 1-800-428-2336 if you have questions about the information contained in this letter.
- Initiated
- 2010-10-11
- Root cause
- Other
- Terminated
- 2011-09-13
- Distribution
- Nationwide distribution.
- Recall number
- Z-1823-2011
- Recall status
- Terminated
- CDRH record ID
- 96430
- FDA classified
- 2011-03-25
- Recall event number
- 57455