Z-1823-2024: Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

Recall record Z-1823-2024 for Fisher & Paykel Healthcare, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1823-2024
Company in FDA record
Fisher & Paykel Healthcare, Ltd.
FDA classification date
May 15, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.

Source facts

Action
On 4/2/24, recall notices were sent to biomedical and respiratory department customers informing them to do the following: 1) Quarantine affected devices and work with firm to arrange collection and replacement. 2) Should a patient need to continue to use the device, the patient is instructed to contact their physician. Distributors are asked to facilitate device collection and replacement for their customers: a) If affected product has been further distributed create a list of affected customers for tracking purposes. b) Use the firm provided templates to create recall letters and response forms. Send the recall notice and response form to your customers to whom your firm has distributed affected devices. c) This recall notice should be shared with anyone who needs to be aware both within your organization and forwarded to any organization/ where potentially affected devices have been transferred. 3) Complete and return the response form via email at FieldAction.USA@fphcare.com Firm is removing and replacing affected devices. If you have any questions, please contact firm Customer Care via email at FieldAction.USA@fphcare.com or directly at +1 (800) 446 3908 ext 5003 or +1 (949) 453 4000 ext 5003.
Initiated
2024-04-02
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Recall number
Z-1823-2024
Recall status
Open, Classified
CDRH record ID
207160
FDA classified
2024-05-15
Recall event number
94398

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗