- FDA identifier
- Z-1823-2025
- Company in FDA record
- Maquet Cardiopulmonary Ag
- FDA classification date
- May 27, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
HKH 8820 Wall Holder does not comply with standard DIN EN 1789:20 and has the potential to detach.
Source facts
- Action
- An URGENT MEDICAL DEVICE RECALL (REMOVAL) notification dated April 2025 was mailed to consignees. The notification instructs consignees to identify and quarantine all affected product. Consignees should forward the provided recall notification to any individuals within their organization that utilize the affected device or to the entity it was distributed to. Consignees should complete and return the provided response form by email to recallresponses.qrc@getinge.com. Getinge Customer Service can be contacted at 888-943-8872 option 2 or by email at ACTSalesSupport.US@getinge.com to coordinate return of affected devices.
- Initiated
- 2025-04-30
- Root cause
- Under Investigation by firm
- Distribution
- Domestic: IL & MO. Foreign: Australia, Austria, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, South Korea, Poland, Russia, Saudi Arabia, Spain, Switzerland, Thailand.
- Recall number
- Z-1823-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213724
- FDA classified
- 2025-05-27
- Recall event number
- 96820