Skip to content

Public FDA record

Z-1824-2011: Calcium Reagent, Calcium Test System for cobas c 111, Part number 04718933190, Roche Diagnostics Co

Recall record Z-1824-2011 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1824-2011
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA classification date
Mar 25, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Internal investigations have confirmed imprecision and erroneous patient results (up to 35% discrepant results) for specific lots of the Calcium assay. Internal investigations revealed the presence of precipitate in the R2 of the affected calcium reagent lots. The precipitate observed was classified as o-cresolphthalein complexone (o-CPC), a reactive component of the calcium R2. The root cau

Source facts

Action
The firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated October 11, 2010 to all consignees/customers via UPS(signature required) . The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the recalled product; discard from inventory according to their site's local regulations; complete and return the attached Faxback form via fax to 1-877-598-5705 even if they do not have the affected product, and file this letter for future reference. The firm added lot 63020401 to the recall in 3/2011. The firm is mailing an URGENT MEDICAL DEVICE REMOVAL notice to customers that purchased that lot of product. Please contact Roche Diagnostics Technical Support at 1-800-428-2336 if you have questions about the information contained in this letter.
Initiated
2010-10-11
Root cause
Other
Terminated
2011-09-13
Distribution
Nationwide distribution.
Recall number
Z-1824-2011
Recall status
Terminated
CDRH record ID
96431
FDA classified
2011-03-25
Recall event number
57455

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗