Z-1824-2014: Spacelabs Healthcare Ventilator Flexport Interface, Model 90436A-07

Recall record Z-1824-2014 for Spacelabs Healthcare Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1824-2014
Company in FDA record
Spacelabs Healthcare Inc
FDA classification date
Jun 18, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Spacelabs Healthcare is voluntarily recalling the Hamilton Galileo Ventilator Flexport, Model 90436A-07, where the monitored Minute Volumes (Vmin) has been reported at one time to reach ten times the actual value on the bedside monitor.

Source facts

Action
Spacelabs changed their strategy for updating the software and sent the updated letter Urgent-Medical Device Correction- Hamilton Galileo Ventilator Flexport Model 90436A-07, dated 10 June 2015, to their consignees. The firm is going to send consignees new replacement flexport(s) with upgraded software and requested consignees to return old affected units via Return Goods Authorization (RGA) to Spacelabs and the firm will destroy affected units. Consignees are instructed to take the following actions: 1. Please replace the affected devices and test for correct operation. 2. Return your old device(s) using the RGA number and prepaid FedX shipping label. Use the Reference code: and FedEx Return number: RETURN ADDRESS: Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 (425) 363-5816 If you have any questions about this corrective action program, please call Spacelabs Healthcare at (800) 522-7025 and select 2 for Technical Support. ********************************************************************************************** Spacelabs recommends that customers immediately advise their staff of this situation and if they continue to use the Flexport, to please keep in mind: 1) Galileo Interface Protocol (GIP) version 1.0 and 1.1 do not affect the flexport readings. Only GIP version 1.2 is an issue. 2) If you have the GIP version 1.2: -The ventilator alarms continue to work normally at the bedside and central station. -The ventilator itself is available and working correctly. -The reported x10 volumes would be an obvious error to the clinician. Spacelabs Field Service personnel will be contacting all customers to schedule a convenient time to install, at no cost, a software update that resolves this issue.
Initiated
2014-05-09
Root cause
Component change control
Terminated
2015-11-16
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, IN, LA, GA, SC, FL, and NY, and 18 customers outside the US.
Recall number
Z-1824-2014
Recall status
Terminated
CDRH record ID
127487
FDA classified
2014-06-18
Recall event number
68374

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗