Small sized specula were incorrectly packaged in printed bags for the medium sized speculum
Source facts
Action
CooperSurgical sent an Urget Medical Device Recall letter dated April 13, 2015, to all affected customers. The letter identified the product, the problem, and the actions to be taken by the customer. Customers were instructed to discontinue use, discard, or return the affected to product to Cooper Surgical. Customers were asked to complete the attached Acknowledgment and Receipt Form for replacement. Customers with questions were instructed to call 203-601-5200.
Initiated
2015-04-13
Root cause
Packaging change control
Terminated
2016-02-05
Distribution
Nationwide Distribution including IL, FL, GA, MA, MO, TX, and WA.
Recall number
Z-1824-2015
Recall status
Terminated
CDRH record ID
137298
FDA classified
2015-06-22
Recall event number
71313