Open original FDA record ↗Recall record
- FDA identifier
- Z-1824-2017
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Mar 30, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until the next reactivation. Following reactivation, the system no longer initializes as intended.
Source facts
- Action
- Siemens mailed a Customer Safety Advisory Notice (CSAN) AX017/17/S to affected customers on 03/28/2017 to inform them of the corrective action that will be performed to prevent a possible hazard to persons and equipment. The notice also explained the underlying issue, the impact on system operation, potential risk to patients and actions they are taking to correct the issue.
- Initiated
- 2017-03-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-08-07
- Distribution
- Distributed to: VA, CA, WI
- Recall number
- Z-1824-2017
- Recall status
- Terminated
- CDRH record ID
- 154432
- FDA classified
- 2017-03-30
- Recall event number
- 76869