Z-1825-2023: Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020

Recall record Z-1825-2023 for Remote Diagnostic Technologies Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1825-2023
Company in FDA record
Remote Diagnostic Technologies Ltd.
FDA classification date
Jun 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".

Source facts

Action
On 5/8/23, Philips mailed correction notices to customers who were asked to do the following: Potential risks associated with partial or complete failure of the device, should a patient be involved, will be mitigated by immediately starting CPR/ chest compressions in case of device dysfunction/failure. Philips will contact to arrange for the device to be returned for rework. A loner device will be provided during the time of the rework. This notice needs to be shared with those who need to be aware within your organization and any organization where the potentially affected devices have been transferred. Complete and return the Customer/Distributor Reply Form: https://philips.efmfeedback.com/se/705E3ED8496AFDF8/ https://philips.efmfeedback.com/se/705E3ED8496AFDF8 Customers/Distributors who need further information or support can contact RDT.Recall.Response@Philips.com, or 1 800 722 9377 In addition, distributors were asked to forward the correction notice to their customers. The notice should be modified to include the distributor's contact information.
Initiated
2023-05-08
Root cause
Device Design
Distribution
US Nationwide distribution in the states of UT, CO, CA, NE, IL, TX, WV, KS, GA.
Recall number
Z-1825-2023
Recall status
Open, Classified
CDRH record ID
199999
FDA classified
2023-06-12
Recall event number
92251

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗