Z-1825-2024: The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitorin

Recall record Z-1825-2024 for Sophysa S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1825-2024
Company in FDA record
Sophysa S.A.
FDA classification date
May 15, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

Source facts

Action
Sophysa issued Urgent Product Recall Letter dated 4/9/24 via email. Letter states reason for recall, health risk and action to take: 1- Please stop using the listed products immediately and remove all affected units from your inventory, 2- Please inform the healthcare professionals at your facility who use these products, 3- Isolate the units concerned in a safe place for return to Sophysa 4- Please complete the enclosed Acknowledgement Form, even if you are no longer in possession of the products to be returned. And return it to Sophysa contact@sophysa.us no later than April 15, 2024. 5- If you have products to return, please pack them in an appropriate shipping box. Upon receipt of the verification form, Sophysa will contact you to arrange the return of the products
Initiated
2024-04-09
Root cause
Process change control
Distribution
Worldwide - US Nationwide distribution in the states of AZ, NY, TX and the countries of ARGENTINA, CHINA, CZECH REPUBLIC, FRANCE, GREECE, HUNGARY, JAPAN, PAKISTAN, Portugal, ROMANIA, SOUTH AFRICA, SWEDEN, VIETNAM.
Recall number
Z-1825-2024
Recall status
Open, Classified
CDRH record ID
207255
FDA classified
2024-05-15
Recall event number
94464

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗