Open original FDA record ↗Recall record
- FDA identifier
- Z-1826-2011
- Company in FDA record
- GE Inspection Technologies, LP
- FDA classification date
- Apr 6, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.
Source facts
- Action
- The firm has completed all of the following necessary actions and no further actions have been indicated: A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
- Initiated
- 2009-03-16
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-04-17
- Distribution
- Nationwide Distribution
- Recall number
- Z-1826-2011
- Recall status
- Terminated
- CDRH record ID
- 82488
- FDA classified
- 2011-04-06
- Recall event number
- 52193