Z-1826-2026: WHILL Model C2, with software versions before HMI: 2.22 / MC: 1.13; device is an indoor/outdoor battery-o

Recall record Z-1826-2026 for WHILL, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1826-2026
Company in FDA record
WHILL, INC.
FDA classification date
Apr 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Cyber vulnerability of BLE communication was found by CISA VDP program.

Source facts

Action
Whill, Inc. issued a Medical Device Recall notice to its sole consignee on 3/23/2026 via email. The notice explained the issue, risks to health, and how to recognize the failure of the device. The notice also provided the following instructions for the distributor: Actions to be taken by the Distributor: 1. Cease all distribution of the devices until the corrective firmware update (2.24 for Model C2 and 2.25 for Model F) has been successfully installed. 2. Notify all customers with previous versions of Model C2 and Model F firmware of this issue. You may use this notice with the User Response Form. 3. Complete the attached response form and return it to XXX." The actions to be taken by the user include explicit instructions on how to disable the Bluetooth so the device may be safely operated. For questions, contact: Email: support.us@whill.inc or Phone: 844-699-4455. The firm has also posted the notice on their website: https://whill.inc/us/voluntary-safety-notice/ On 6/29/2026 Whill, Inc. issued an updated notice to its consignees, distributor and user, informing them of the availability of the updated firmware to correct the device with detailed instructions on installation. The firm also placed a copy of the notice on its website.
Initiated
2025-12-29
Root cause
Device Design
Distribution
US: Kansas
Recall number
Z-1826-2026
Recall status
Open, Classified
CDRH record ID
218240
FDA classified
2026-04-24
Recall event number
98320

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗