Skip to content

Public FDA record

Z-1827-2020: CapsoCAM Plus, UDI: 00867770000209

Recall record Z-1827-2020 for Capso Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1827-2020
Company in FDA record
Capso Vision, Inc.
FDA record creation date
Apr 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Capsule Endoscopy System: a defect present in the capsule window may be stressed through the production process and in the final package, which could cause the housing of the capsule to crack and leak when ingested. Fluid from the patient could leak in to the capsule, which may damage components inside the capsule and the patient data may not be retrieved; the patient may need to repeat the exam.

Source facts

Action
Starting on 10/18/19, Urgent Medical Device Voluntary Recall Letters were emailed to customers. Customers were asked return affected product and to complete the Return Material Authorization form. The recalling firm asked distributors to provide their latest shipment lists.
Initiated
2019-10-18
Root cause
Other
Terminated
2021-01-08
Distribution
U.S.: CA, TX, FL, VA, PA, SC, HI, DC, NJ, UT, LA. Foreign: FR, AR, IT, UK, BE, TC, AU, CO, GM, GR, SP, SI, HR, MO, SW, EZ.
Recall number
Z-1827-2020
Recall status
Terminated
CDRH record ID
180717
FDA record created
2020-04-30
Recall event number
85378

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗