Open original FDA record ↗Recall record
- FDA identifier
- Z-1827-2024
- Company in FDA record
- Philips North America Llc
- FDA classification date
- May 15, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Event Catalog information does not save when copied and transferred from one unit to another.
Source facts
- Action
- Consignees were sent an "URGENT Medical Device Correction," notification, dated 4/12/24. Consignees are instructed in the notification to review in Events Monitor and confirm the expected category is assigned for the reviewed event. If events are not assigned to the correct category consignees are not to depend on event notification for patient monitoring. The provided recall notification should be provided to all users of affected devices and to those the device was further distributed, for their awareness. Philips will contact consignees to schedule a visit from a Philips Field Service Engineer to provide a software patch (PIC iX 4.2.4) to correct the issue. Consignees with any questions are to contact their local Philips representative or call 1-800-722-9377.
- Initiated
- 2024-04-19
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, MA, MI, MN, MO, NJ, NM, NY, OH, PA, TN, TX, WA, WV, & DC. The countries of AE, AT, AU, BE, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, IE, IL, IN, IT, JP, KR, LT, NL, NO, OM, PL, PT, RO, SE, SG, SK, TH, & ZA.
- Recall number
- Z-1827-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207266
- FDA classified
- 2024-05-15
- Recall event number
- 94468