Open original FDA record ↗Recall record
- FDA identifier
- Z-1828-2008
- Company in FDA record
- Mallinckrodt Inc
- FDA classification date
- Sep 3, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The tubing metal begins to crack and thin. When this occurs it causes the tubing material to crack around the screws. This can cause the tubing to fall from the mounting plug.
Source facts
- Action
- Letter dated April 18, 2008. The firm has requested their customers check their inventory, and determine if they have any of the affected suspension systems. If they observe any loose screws or cracking around the J-Bow tube, immediately take the system out of service. A service representative will contact them to schedule any required system corrections. For assistance, contact Product Monitoring Department at 1-800-778-7898 or by email at productmonitoringhazelwood@covidien.com.
- Initiated
- 2008-04-14
- Root cause
- Device Design
- Terminated
- 2011-05-20
- Distribution
- Product was sold in 46 US states: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV & WY,
- Recall number
- Z-1828-2008
- Recall status
- Terminated
- CDRH record ID
- 70474
- FDA classified
- 2008-09-03
- Recall event number
- 47889