Skip to content

Public FDA record

Z-1828-2008: Covidien/ Mallinckrodt CT9000, CT9000 ADV (J-Bow component 241300), contrast media delivery system for d

Recall record Z-1828-2008 for Mallinckrodt Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1828-2008
Company in FDA record
Mallinckrodt Inc
FDA classification date
Sep 3, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The tubing metal begins to crack and thin. When this occurs it causes the tubing material to crack around the screws. This can cause the tubing to fall from the mounting plug.

Source facts

Action
Letter dated April 18, 2008. The firm has requested their customers check their inventory, and determine if they have any of the affected suspension systems. If they observe any loose screws or cracking around the J-Bow tube, immediately take the system out of service. A service representative will contact them to schedule any required system corrections. For assistance, contact Product Monitoring Department at 1-800-778-7898 or by email at productmonitoringhazelwood@covidien.com.
Initiated
2008-04-14
Root cause
Device Design
Terminated
2011-05-20
Distribution
Product was sold in 46 US states: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV & WY,
Recall number
Z-1828-2008
Recall status
Terminated
CDRH record ID
70474
FDA classified
2008-09-03
Recall event number
47889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗