Z-1828-2011: Phoenix x/ray nanotom series. Phoenix x|ray nanotom series cabinet x-ray systems are used in non-destruc

Recall record Z-1828-2011 for GE Inspection Technologies, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1828-2011
Company in FDA record
GE Inspection Technologies, LP
FDA classification date
Apr 6, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.

Source facts

Action
The firm has completed all of the following necessary actions and no further actions have been indicated: A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
Initiated
2009-03-16
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-04-17
Distribution
Nationwide Distribution
Recall number
Z-1828-2011
Recall status
Terminated
CDRH record ID
82490
FDA classified
2011-04-06
Recall event number
52193

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗