Open original FDA record ↗Recall record
- FDA identifier
- Z-1828-2015
- Company in FDA record
- Terumo Cardiovascular Systems Corp
- FDA classification date
- Jun 23, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.
Source facts
- Action
- Consignees were notified of the recall by phone beginning on 5/7/2015 and a Urgent Medical Device Recall letter, dated 5/7/2015. The letter identified the affected device, reason for recall, potential hazard, and correction. Customers were asked to review the notice and ensure that all users are aware of the issue. In addition, they are to confirm the quantity of procedure kits for return; and, complete and return the Customer Response Form. Customers can call Terumo CVS Customer Service at 1-800-521-2818 with questions.
- Initiated
- 2015-05-06
- Root cause
- Component design/selection
- Terminated
- 2015-07-27
- Distribution
- Distributed to the states of MO and FL.
- Recall number
- Z-1828-2015
- Recall status
- Terminated
- CDRH record ID
- 136456
- FDA classified
- 2015-06-23
- Recall event number
- 71196