Z-1829-2008: GE Datex-Ohmeda Engstrom Carestation The Carestation consists of three main components: a display, a vent

Recall record Z-1829-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1829-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 24, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a correction associated with Engstrom Carestation with an integrated Aerogen Professional Nebulizer System. There is an issue relating to the removal of the nebulizer cable from nebulizer connection port, which, when done while the nebulizer is active, may result in the discontinuance of mechanical ventilation. This issue affects all of the Engstrm Carestations shipped between 12/4/2006

Source facts

Action
Consignees were sent a "GE Urgent Medical Device Correction" letter on 5/12/08 addressed to Health Care Administrator/Risk Manger, Chief of Intensive Care & Director of Clinical Engineering. The letter described the affected product and the problem and provided a short term recommendation and a long term solution which included installation of a revised board on all affected Engstrom Carestations. A GE Healthcare Field Service Engineer to contact the facility to schedule the update. If you have questions, contact Melissa Davis at GE Healthcare at 1-800-345-2700 extension 3439.
Initiated
2008-04-23
Root cause
Device Design
Terminated
2012-05-16
Distribution
WI, GA, SC, TX, RI, NY, NM, IN VIETNAM, VENEZULA, UNITED KINGDOM, TURKEY, KOREA, JAPAN, ITALY, INDIA, ICELAND, HONG KONG, GREECE, GERMANY, GEORGIA, FRANCE, TUNISIA, THAILAND, SWITZERLAND, SPAIN, SOUTH AFRICA, FINLAND, ESTONIA, DENMARK, CZECH REPUBLIC, CROATIA, COLOMBIA, CHINA, CANADA, BRAZIL, BELGIUM, BANGLADESH, BAHRAIN, SLOVENIA, SINGAPORE, SAUDI, RUSSIA, QATAR, PORTUGAL, POLAND, PHILIPPINES, NETHERLAND,MOROCCO, MEXICO, MAYLASIA, LYBIA, LEBANON, KUWAIT, AUSTRIA, AUSTRALIA.
Recall number
Z-1829-2008
Recall status
Terminated
CDRH record ID
70814
FDA classified
2008-09-24
Recall event number
48080

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗