Open original FDA record ↗Recall record
- FDA identifier
- Z-1829-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Jun 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement
Source facts
- Action
- Philips Healthcare Imaging issued Urgent Field Safety Notice dated May 21, 2012, to accounts informing an inspection will be performed of the unit. The Cable guiding plate will be replaced. If the cables show any damage the complete cable set wll also be replaced. No action is be taken by the customer. For further information customers should call 1-800-722-9377. For questions regarding this recall call 978-687-1501.
- Initiated
- 2012-05-21
- Root cause
- Device Design
- Terminated
- 2017-01-18
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Australia Austria Belgium Bharain Canada China Columbia Czech Republic Denmark Dutch Antilles Egypt Estonia Finland France Germany Hong Kong Ireland Israel Italy Latvia Lebanon Luxembourg Malaysia Mexico Netherlands New Zealand Norway Philippines Qatar Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syria Taiwan Thailand Turkey United Kingdom
- Recall number
- Z-1829-2012
- Recall status
- Terminated
- CDRH record ID
- 109739
- FDA classified
- 2012-06-20
- Recall event number
- 62052