Skip to content

Public FDA record

Z-1829-2012: Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva w

Recall record Z-1829-2012 for Philips Healthcare Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1829-2012
Company in FDA record
Philips Healthcare Inc.
FDA classification date
Jun 20, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement

Source facts

Action
Philips Healthcare Imaging issued Urgent Field Safety Notice dated May 21, 2012, to accounts informing an inspection will be performed of the unit. The Cable guiding plate will be replaced. If the cables show any damage the complete cable set wll also be replaced. No action is be taken by the customer. For further information customers should call 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Initiated
2012-05-21
Root cause
Device Design
Terminated
2017-01-18
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Australia Austria Belgium Bharain Canada China Columbia Czech Republic Denmark Dutch Antilles Egypt Estonia Finland France Germany Hong Kong Ireland Israel Italy Latvia Lebanon Luxembourg Malaysia Mexico Netherlands New Zealand Norway Philippines Qatar Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syria Taiwan Thailand Turkey United Kingdom
Recall number
Z-1829-2012
Recall status
Terminated
CDRH record ID
109739
FDA classified
2012-06-20
Recall event number
62052

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗