Open original FDA record ↗Recall record
- FDA identifier
- Z-1829-2015
- Company in FDA record
- Varian Medical Systems Inc
- FDA classification date
- Jun 23, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06. The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat
Source facts
- Action
- All the consignees affected by this recall will be notified by: The Urgent Medical Device Correction Letter/ Field Safety Notification will be distributed to all affected users, with a description of the problem. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. Additionally, the letter will be posted to the customer support site: http:/ /www.MyVarian.com.
- Initiated
- 2015-05-18
- Root cause
- Process design
- Terminated
- 2016-01-26
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1829-2015
- Recall status
- Terminated
- CDRH record ID
- 137839
- FDA classified
- 2015-06-23
- Recall event number
- 71450