Z-1830-2015: LapSac Surgical Tissue Pouch. The current intended use for the LapSac Surgical Tissue Pouch is for isol

Recall record Z-1830-2015 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1830-2015
Company in FDA record
Cook Inc.
FDA classification date
Jun 24, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Includes Information not stated in the intended use.

Source facts

Action
Cook Medical sent an "Recall Communication Updated Instructions for Use" letter dated May 15, 2015, to all affected customers . The letter identified the product the problem and the action needed to be taken by the customer. Cook has requested that customers do the following: The intended use for the device is being modified from "Used to isolate tissue" during, or prior to, surgical removal and/or morcellation to "Used to isolate tissue" during, or prior to, surgical removal. A contraindication being added is as follows: " Not for use with power morcellators. The warnings are being modified to read as follows: " The LapSac Surgical Tissue Pouch is not intended for use with sharp surgical instruments. " Care should be taken to avoid damaging, puncturing or tearing the LapSac Surgical Tissue Pouch. " Use caution when removing potentially malignant or harmful tissues, to isolate contamination from surrounding tissue. The instructions for use are being modified to read as follows: 1. Following aseptic technique, introduce the LapSac Surgical Tissue Pouch into the abdominal cavity. 2. Place tissue to be removed from the body into the pouch. 3. Using the drawstring, carefully close the LapSac Surgical Tissue Pouch. Please note that if excess force is used, the drawstring may break or tear the LapSac. 4. Remove the LapSac Surgical Tissue Pouch from the body, taking care not to spill contents of the bag. Please report any Adverse Event to Cook Medical Customer Relations (800) 457-4500 or 1-812-339-2235. Monday through Friday between 7:30 a.m. and 5:00 p.m. Eastern Daylight Time or email at CustomerrelationsNA@cookmedical.com; and to FDA. The customer notice must be passed on to appropriate personnel within each organization or to any organization where the potentially affected devices have been transferred. If customers have further questions regarding labelling changes for this device, please contact Cook Sales Representative or Cook M
Initiated
2015-05-14
Root cause
Labeling Change Control
Terminated
2016-09-12
Distribution
Worldwide Distribution - US (nationwide) including PR and to the states of: AZ, AK, AL, AR, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, MO, ME, MT, NC, ND, NE, NH, NJ, NY, NM, OH,OK,OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY., and to the countries of : China, Korea, Denmark, Mexico, New Zealand, India and Colombia.
Recall number
Z-1830-2015
Recall status
Terminated
CDRH record ID
137267
FDA classified
2015-06-24
Recall event number
71287

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗