In specific circumstances, the user is able to restart an aborted irradiation without having to select the correct partial irradiation data.
Source facts
Action
A Field Safety Notice was mailed to the customer affected by the issue on 03/21/2017. The Notice informed the user when resuming from a partial irradiation, to ensure that the correct delivered dose has been selected in the partial report dialog box before proceeding with irradiation.
Initiated
2017-03-21
Root cause
Software design
Terminated
2018-09-19
Distribution
Distribution in US (to Louisiana), and France.
Recall number
Z-1830-2017
Recall status
Terminated
CDRH record ID
154230
FDA record created
2017-04-17
Recall event number
76804