Skip to content

Public FDA record

Z-1831-2017: NEODENT CM Intraoral Scanbody, Article Number: 108.139

Recall record Z-1831-2017 for Instradent USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1831-2017
Company in FDA record
Instradent USA, Inc.
FDA classification date
Apr 7, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During internal testing, Neodent determined that the body diameter (3.8mm) of the Neodent 108.139 CM Exact Intraoral Scan body is not compatible with the platform diameter of CM Implants when the implant is intraosseoussly (subcrestally) positioned.

Source facts

Action
Instradent sent an Urgent Field Safety Notice dated March 13, 2017, to all affected customers via UPS. Customers were asked to take the following actions: 1. Identify and segregate the affected units from your stock. 2. If the product is still in your Inventory, return it for credit. 3. If you have used the product successfully, there is no need to take any action 4. For all cases complete and return the enclosed Customer Confirmation Form. For further questions, please call (855) 412-8883.
Initiated
2017-03-13
Root cause
Component design/selection
Terminated
2019-02-28
Distribution
Distributed to the states of: NJ, TX, NV, CA, PA, AZ, and IL.
Recall number
Z-1831-2017
Recall status
Terminated
CDRH record ID
154385
FDA classified
2017-04-07
Recall event number
76843

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗