Open original FDA record ↗Recall record
- FDA identifier
- Z-1831-2017
- Company in FDA record
- Instradent USA, Inc.
- FDA classification date
- Apr 7, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During internal testing, Neodent determined that the body diameter (3.8mm) of the Neodent 108.139 CM Exact Intraoral Scan body is not compatible with the platform diameter of CM Implants when the implant is intraosseoussly (subcrestally) positioned.
Source facts
- Action
- Instradent sent an Urgent Field Safety Notice dated March 13, 2017, to all affected customers via UPS. Customers were asked to take the following actions: 1. Identify and segregate the affected units from your stock. 2. If the product is still in your Inventory, return it for credit. 3. If you have used the product successfully, there is no need to take any action 4. For all cases complete and return the enclosed Customer Confirmation Form. For further questions, please call (855) 412-8883.
- Initiated
- 2017-03-13
- Root cause
- Component design/selection
- Terminated
- 2019-02-28
- Distribution
- Distributed to the states of: NJ, TX, NV, CA, PA, AZ, and IL.
- Recall number
- Z-1831-2017
- Recall status
- Terminated
- CDRH record ID
- 154385
- FDA classified
- 2017-04-07
- Recall event number
- 76843