Z-1831-2024: CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

Recall record Z-1831-2024 for Maquet Medical Systems USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1831-2024
Company in FDA record
Maquet Medical Systems USA
FDA classification date
May 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.

Source facts

Action
Consignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can also be contacted at 888-943-8872 (option 4, option 2, option 1) Monday through Friday between 8:00 AM and 6:00 PM EST to make service arrangements. Consignees are asked to not use affected devices for any additional patients until it has been cleared by the MCP/Getinge system check, repaired, or replaced. The provided recall notice is to be forwarded to users of the CARDIOHELP-i system and if the device was further distributed.
Initiated
2024-04-15
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of NJ & TX.
Recall number
Z-1831-2024
Recall status
Open, Classified
CDRH record ID
207234
FDA classified
2024-05-16
Recall event number
94450

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗