Open original FDA record ↗Recall record
- FDA identifier
- Z-1832-2013
- Company in FDA record
- Exactech, Inc.
- FDA classification date
- Jul 31, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument.
Source facts
- Action
- Exactech sent a" Important Product Recall Notice" dated June 4, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were requested to Immediately cease distribution or use of the products. Extend this information to your accounts that may have this product in their possession. Verify if you have any of the subject ARTICULATE DRIVER (catalog number 181-00-04) in the specified lots. FAx back the attached form. Further questions please call 1-800-392-2832.
- Initiated
- 2013-05-30
- Root cause
- Under Investigation by firm
- Terminated
- 2013-09-19
- Distribution
- US Distribution including the states of AR, CA, FL, NJ, NY, OK, and TX
- Recall number
- Z-1832-2013
- Recall status
- Terminated
- CDRH record ID
- 119601
- FDA classified
- 2013-07-31
- Recall event number
- 65568