Z-1832-2013: Exactech Articulated Driver A nonpowered hand-held device intended for medical purposes to manipulate t

Recall record Z-1832-2013 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1832-2013
Company in FDA record
Exactech, Inc.
FDA classification date
Jul 31, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument.

Source facts

Action
Exactech sent a" Important Product Recall Notice" dated June 4, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were requested to Immediately cease distribution or use of the products. Extend this information to your accounts that may have this product in their possession. Verify if you have any of the subject ARTICULATE DRIVER (catalog number 181-00-04) in the specified lots. FAx back the attached form. Further questions please call 1-800-392-2832.
Initiated
2013-05-30
Root cause
Under Investigation by firm
Terminated
2013-09-19
Distribution
US Distribution including the states of AR, CA, FL, NJ, NY, OK, and TX
Recall number
Z-1832-2013
Recall status
Terminated
CDRH record ID
119601
FDA classified
2013-07-31
Recall event number
65568

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗