Open original FDA record ↗Recall record
- FDA identifier
- Z-1832-2024
- Company in FDA record
- Cook Medical Incorporated
- FDA classification date
- May 16, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Devices may contain elevated levels of bacterial endotoxin.
Source facts
- Action
- On April 01, 2024, Cook Ireland Ltd.(CIRL) initiated a field action for the EchoTip Ultra High Definition Ultrasound Access Needle (ECHO-HD-19-A) and EchoTip Ultra Endoscopic Ultrasound Needle (ECHO-19). Cook Ireland Ltd has identified 9 devices located outside of the United States (US) that are associated with the issue and has initiated a voluntary recall due to the potential of elevated levels of bacterial endotoxins on the devices. No US customers were impacted by this Field Safety Corrective Action (FSCA). No monitoring will be required as US consignees are not impacted by this recall. Effectiveness checks for this recall will not be necessary as US Consignees were not impacted by the distribution.
- Initiated
- 2024-04-01
- Root cause
- Nonconforming Material/Component
- Distribution
- International distribution in the countries of Germany, Netherlands, Switzerland, United Kingdom.
- Recall number
- Z-1832-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207156
- FDA classified
- 2024-05-16
- Recall event number
- 94395