A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose, for the OAR (Organs at Risk) the user only defines the max. dose. The firms engineering team determined that the software defect was repeatable.
Source facts
Action
Affected consignees will be notified by the Customer Service Engineer to schedule a software update beginning November 2009.
Initiated
2009-11-06
Root cause
Software design
Terminated
2011-04-06
Distribution
Worldwide distribution: USA including: CA, KY, MI, NJ and TN; and countries of: Canada and Mexico.
Recall number
Z-1834-2011
Recall status
Terminated
CDRH record ID
89038
FDA classified
2011-03-28
Recall event number
54706