Open original FDA record ↗Recall record
- FDA identifier
- Z-1834-2013
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Jul 31, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash solution pump to stop, creating an overflow or leak on each analyzer unit.
Source facts
- Action
- An Urgent Product Correction letter dated 5/29/13 was sent on the week of June 3, 2013 to all the customers who received the AU5800 Clinical Chemistry Analyzer with Automation Ready (AR) Connection (All Software versions). The letter informed the customers of the problem identified and the actions to be taken. Customers were instructed to complete and return the enclosed Response Form within 10 days and to share the information with their laboratory staff. Customers with questions are instructed to contact the 24 hours Call Center (Hotline) at (800) 854-3633 in the US and Canada.
- Initiated
- 2013-06-03
- Root cause
- Software design
- Terminated
- 2014-08-23
- Distribution
- Worldwide Distribution, including Nationwide (US) and the countries of China, Italy, Japan, Canada, Turkey, United Kingdom, Singapore, Israel, and the Netherlands.
- Recall number
- Z-1834-2013
- Recall status
- Terminated
- CDRH record ID
- 118910
- FDA classified
- 2013-07-31
- Recall event number
- 65440