Open original FDA record ↗Recall record
- FDA identifier
- Z-1834-2017
- Company in FDA record
- Signal Medical Corporation
- FDA classification date
- Apr 17, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for poly tip of the mallet to come off during use. Potential for foreign material in the surgical field or delay in surgery. Reminder for customers that the poly tip must be removed for cleaning and sterilization to assure proper cleaning and sterilization of the entire mallet.
Source facts
- Action
- On 2/20/2016, URGENT: MEDICAL DEVICE RECALL notifications dated February 10, 2017 were sent to the affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers are to retain the notice and replace the old Instructions For Use with the updated version. Also, they are to acknowledge receipt of the instructions by signing the notice and emailing to SignalMedical2017@gmail.com or faxing to 810-364-7072. The firm has attached the revised labeling to the letter.
- Initiated
- 2017-02-20
- Root cause
- Labeling design
- Terminated
- 2017-08-25
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1834-2017
- Recall status
- Terminated
- CDRH record ID
- 153400
- FDA classified
- 2017-04-17
- Recall event number
- 76509