Open original FDA record ↗Recall record
- FDA identifier
- Z-1835-2011
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Mar 28, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Stryker Orthopaedics has identified a trend of complaints involving reports of the peg drill cold welding to the drill guide, causing the Peg Drill to cease functioning. If cold welding of these components occurs, the following potential patient and/or surgeon or surgical staff harms may results.
Source facts
- Action
- Stryker sent Urgent Product Recall letters on February 9, 2011, and February 10, 2011, via Fed Ex. to all affected customers. The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete and fax back the attached Product Recall Acknowledgement Form as soon as possible to 201-831-6069. For questions regarding this recall call 201-972-2100.
- Initiated
- 2011-02-09
- Root cause
- Component change control
- Terminated
- 2013-11-18
- Distribution
- Worldwide Distribution - USA , Sweden, Germany; France, Portugal, UK, Canada, Australia, and New Zealand,
- Recall number
- Z-1835-2011
- Recall status
- Terminated
- CDRH record ID
- 97590
- FDA classified
- 2011-03-28
- Recall event number
- 57857